Discover elegant solution for Your Business
multi channel Syringe Pump LC1106
Model No.︰
LC-1106
Brand Name︰
Landcare
Country of Origin︰
China
Inquire Now
Total 6 Related Items
Product Description
Main Features for LC-1106 Syringe Pump:
Antibolus feature
KVO(Keep Vein Open)
Near-end alarm to meet clinic needs
With bolus rate(purge rate) function
Adopt double MPU to ensure system security
Occlusion alarm pressure is selectable
Automatically recognize the syringe type
Temporarily clear up alarm function and timing to recover
Steamless-convert power supply system with internal battery embeded
Pressing on/off button for 3 seconds is needed to turn off to avoid mistaken operation
No changing the data during operation to avoid mistaken operation.
Detailed Spec.
Model No.
LC-1106
Channel
Four Channel
Syringe Selection
10 ml, 20ml, 30ml,50/60 ml
Highest Infusion Rate
400 ml/h
Infusion Rate
0.1~400ml/h(50ml/60ml)
0.1~200ml/h(20ml/30ml)
0.1~99.9ml(10ml)
Purge/Bolus Rate
400ml/h(50ml/60ml)
200ml/h(20ml/30ml)
99.9ml(10ml)
Infusion Volume
0~999.9ml, step by 0.1
Display Volume
0~999.9ml
Infusion Rtae Step
0.1ml/h
Accuracy
±3%
Infusion Rate Step
0.1ml/h
Alarms
Syringe installation failure,Occlcusion,Near end,Pause overtime,Low-battery,Completion, Power interruption, Motor malfunction, Incorrect specification of the syringe,Electrical circuit jam.
Occlusion Pressure:High
20 ml,30ml,50/60ml syringe spec:90 Kpa±30Kpa; 10 ml syringe spec:100Kpa±30Kpa
Occlusion Pressure:Low
20 ml,30ml,50/60ml syringe spec:50Kpa±30Kpa; 10 ml syringe spec:70Kpa±30Kpa
KVO Rate
0.1ml/h±0.02ml/h
Pause
AC:100V~240V 50Hz/60Hz DC:12V
Pause Overtime Alarm
1 min 50s~2min
Payment Terms︰
TT / LC
Product Image