Model No.︰LC-1003
Brand Name︰Landcare
Country of Origin︰China
Unit Price︰-
Minimum Order︰-
Main Features for LC-1003 Infusion Pump:
1.with bolus rate
2.Available for most brand of I.V sets after an easy calibration
3.with internal battery which can be disassembled for easier transportation.
4.the flow rare in ml/h and drop/min can convert automatically.
5.Double MPU in work to ensure system security
6.With intelligent interface on back panel,can update unit software & connect to nurse call jack
7.Ultrasound detection of air and selectable drop sensor can make the infusion safer.
Model No. | LC-1003 | |
Range of pre-set of infusion rate | 1.0~1200ml/h (0.1ml/h step 1.0~99.9ml/h;1ml/h step 100~1000ml/h) | |
Bolus rate | 800~1200ml/h(default value is 800ml/h) | |
Range of pre-set of infusion volume | 1-9999ml | |
Indication range of volume infused | 1-9999ml | |
Use of syringe | Comply with all brands after calibration | |
Infusion precise | ±5% | |
Bubble sensitivity | Detestability of air bubble ≥40uL | |
Occlusion alarm pressure | 100KPa±20KPa | |
Alarming | Air Bubble, Occlusion, Door Open, Complete infusion, Low power, Motor failure,Drop abnormal, Low temperature,Pause overtime ,Electrical circuit jam,Error set of accuracy,Air detect circuit error,Motor control error,Battery exhausted | |
KVO rate | KVO rate as at 1ml/h with flow rate no higher than 10ml/h; as at 3 ml/h with flow rate higher than10ml/h | |
Internal battery | Fully charged 6 hours, the internal DC12V Ni-MH battery can continuously run over 4 hours at average rate level | |
Power | 100-240VAC 50/60Hz | |
Power consumption | 18VA | |
Working enviroment | Operation Temperature | 5℃-40℃ |
Atmospheric Pressure | 700hPa-1060hPa | |
Ralative Humidity | 20%-90% |
Payment Terms︰ TT / LC