Main Features for LC-1002 Infusion Pump:
1.with bolus rate
2.Available for most brand of I.V sets after an easy calibration
3.with internal battery which can be disassembled for easier transportation.
4.the flow rare in ml/h and drop/min can convert automatically.
5.Double MPU in work to ensure system security
6.With intelligent interface on back panel,can update unit software & connect to nurse call jack
7.Ultrasound detection of air and selectable drop sensor can make the infusion safer.
Detailed spec.
Model No. | LC-1002 | |
Flow Rate | 1.0~800ml/h(0.1 ml/h step 100~800ml/h) | |
Bolus Rate | 600~800 ml/h(default value is 600ml/h) | |
Volume Limit | 1-9999ml | |
Applicable IV sets | Comply with all brands after calibration | |
Accuracy | ±5% | |
Bubble sensitivity | Detestability of air bubble ≥40uL | |
Occlusion alarm pressure | 100Kpa±20KPa | |
Alarming | Air Bubble,Occlusion,Door Open,Complete infusion, Low power,Motor failure,Drop abnormal,Low temprature,Pause overtime,Electrical circuit jam,Error set of accuracy,Air detect circuit error,Motor control error,battery exhausted | |
Drop Sensor | Selectable | |
KVO rate | KVO rate as at 1 ml/h with flow rate no higher than 10 ml/h; as at 3 ml/h with flow rate higher than 10 ml/h | |
Internal Battery | Fully charged 6 hours,the internal DC 12V Ni-MH battery can continously run over 4 hours at average rate level. | |
Power | 100-240V AC 50/60 Hz | |
Power consumption | 18VA | |
Working enviroment | Operation Temperature | 5℃-40℃ |
Atmospheric Pressure | 700hPa-1060hPa | |
Ralative Humidity | 20%-90% |