electron Infusion Pump LC1003

Model No.︰

LC-1003

Brand Name︰

Landcare

Country of Origin︰

China

Unit Price︰

-

Minimum Order︰

-

Share on:

Product Description

Main Features for LC-1003 Infusion Pump:
1.with bolus rate
2.Available for most brand of I.V sets after an easy calibration
3.with internal battery which can be disassembled for easier transportation.
4.the flow rare in ml/h and drop/min can convert automatically.
5.Double MPU in work to ensure system security
6.With intelligent interface on back panel,can update unit software & connect to nurse call jack
7.Ultrasound detection of air and selectable drop sensor can make the infusion safer.

Model No. LC-1003
Range of pre-set of infusion rate 1.0~1200ml/h (0.1ml/h step 1.0~99.9ml/h;1ml/h step 100~1000ml/h)
Bolus rate 800~1200ml/h(default value is 800ml/h)
Range of pre-set of infusion volume 1-9999ml
Indication range of volume infused 1-9999ml
Use of syringe Comply with all brands after calibration
Infusion precise ±5%
Bubble sensitivity Detestability of air bubble ≥40uL
Occlusion alarm pressure 100KPa±20KPa
Alarming Air Bubble, Occlusion, Door Open, Complete infusion, Low power, Motor failure,Drop abnormal, Low temperature,Pause overtime ,Electrical circuit jam,Error set of accuracy,Air detect circuit error,Motor control error,Battery exhausted
KVO rate KVO rate as at 1ml/h with flow rate no higher than 10ml/h; as at 3 ml/h with flow rate higher than10ml/h
Internal battery Fully charged 6 hours, the internal DC12V Ni-MH battery can continuously run over 4 hours at average rate level
Power 100-240VAC 50/60Hz
Power consumption 18VA
Working enviroment Operation Temperature 5℃-40℃
Atmospheric Pressure 700hPa-1060hPa
Ralative Humidity 20%-90%


Payment Terms︰ TT / LC